Passa a contenuto principale
  Tags                  
Languages Lingua - Inglese Lingua - Giapponese Lingua - Tedesco Lingua - Cinese Lingua - Spagnolo Lingua - Italiano Lingua - Francese Lingua - Portoghese    
Language Quality HT MT AT NT INT - Internal          
All Tutto - Prodotti FARO Hardware - Tutti i prodotti FARO                
Quantum FaroArm/ScanArm - Quantum S FaroArm/ScanArm - Quantum M FaroArm/ScanArm - Quantum E              
Faro/Scan Arm FaroArm/ScanArm - Edge FaroArm/ScanArm - Fusion FaroArm/ScanArm - Prime FaroArm/ScanArm - Platinum FaroArm/ScanArm - Legacy Quantum FaroArm/ScanArm - Titanium FaroArm/ScanArm - Advantage FaroArm/ScanArm - Digital Template    
Serial FaroArm Serial FaroArm - Bronze Serial FaroArm - Gold Serial FaroArm - Silver              
Gage Gage - Bluetooth Gage - Plus Gage - Standard Gage - Power            
ScanArm ScanArm - Design ScanArm 2.5C ScanArm - Design ScanArm 2.0 ScanArm - Forensic ScanArm ScanArm - Design ScanArm            
Laser Tracker Laser Tracker - Vantage S6 Laser Tracker - Vantage E6 Laser Tracker - Vantage S Laser Tracker - Vantage E Laser Tracker - Vantage Laser Tracker - ION Laser Tracker - Si Laser Tracker - X Laser Tracker - Xi  
Imager 3D Imager - Cobalt                  
Laser Scanner Scanner laser 3D - Focus S Scanner laser 3D - Focus M Scanner laser 3D - Focus3D Scanner laser 3D - Focus3D X Scanner laser 3D - Focus3D X HDR Scanner laser 3D - Focus3D S Scanner laser 3D - Photon      
Hand Held Scanner Scanner tenuto in mano 3D - Freestyle3D Scanner tenuto in mano 3D - Freestyle3D X Scanner tenuto in mano 3D - Freestyle3D Objects Scanner tenuto in mano 2D - ScanPlan            
Laser Projector Laser Projector - Tracer M Laser Projector - Tracer SI Testa di scansione - Digi-Cube              
Imaging Laser Radar Imaging Laser Radar - VectorRI                  
Computers Computer - Tutti i computer                  
CAM2 CAM2 - CAM2 CAM2 - Measure CAM2 - SmartInspect              
SCENE SCENE - Tutta la suite SCENE SCENE - Acquisizione ed elaborazione SCENE - WebShare Cloud SCENE - WebShare 2Go App SCENE - WebShare Server e Webshare 2Go          
RevEng RevEng - Editor FARO - ScanArm Mesh                
Legacy Software Software legacy - CAM2 Gage Software legacy - Software Gage Software legacy - Insight              
Zone & ARAS FARO CAD Zone - Fire & Insurance FARO CAD Zone - Crime & Crash FARO CAD Zone - CZ Point Cloud FARO CAD Zone - First Look Pro FARO Zone - 2D FARO Zone - 3D FARO 360 - Reality FARO 360 - HD FARO 360 - Blitz FARO 360 - Genius
As-Built As‑Built - AutoCAD As‑Built - Revit VirtuSurv - VirtuSurv 2018              
PointSense1 PointSense - Basic PointSense - Pro PointSense - Building PointSense - Plant PointSense - Heritage PointSense - Revit CAD Plugin - TachyCAD Building CAD Plugin - TachyCAD Archeology CAD Plugin - TachyCAD Interior  
PointSense2 CAD Plugin - PhoToPlan Basic CAD Plugin - PhoToPlan CAD Plugin - PhoToPlan Pro CAD Plugin - PhoToPlan Ultimate CAD Plugin - DisToPlan CAD Plugin - MonuMap CAD Plugin - hylasFM CAD Plugin - VirtuSurv    
VI-Tracer-BuildIT Visual Inspect - App Visual Inspect - CAD Translator RayTracer - RayTracer BuildIT - Metrology BuildIT - Projector BuildIT - Construction        

FaroArm/ScanArm

Quantum S Max

Quantum M Max

Quantum E Max

Gage Max

Quantum S

Quantum M

Gage

FARO® Base di Conoscenza

iProbe non riconosciuto in FaroArm Manager

Panoramica

I problemi legati alla comunicazione difettosa di iProbes con FaroArm® o al mancato riconoscimento di iProbes in FaroArm Manager possono essere dovuti a dati corrotti. Questo articolo spiega passo-passo come cancellare le sonde dal FaroArm Manager in modo da poterle nuovamente installare e compensare.

Requisiti

  • Driver USB 6.10.48 o successivo installato. Scaricare e installare il driver FaroArm più recente.

Procedura dettagliata

  1. Avviare FaroArm Manager e collegarsi a FaroArm con un cavo USB, cliccare su Probe Management (Gestione sonda).
    Deleting corrupted i-probe_Fig.1.jpg
  2. Cliccare sull'icona del cestino per eliminare l'elenco delle sonde e poi cliccare su Yes per confermare l'eliminazione dell'elenco delle sonde.
    Deleting corrupted i-probe_Fig.2.jpg
  3. L'elenco delle sonde deve visualizzare solo la sonda collegata, con una cronologia di calibrazione Never (Mai). Chiudere FaroArm Manager cliccando sulla X.
    Deleting corrupted i-probe_Fig.3.jpg
  4. Spegnere il FaroArm premendo il pulsante sul retro dell'unità, quindi riaccendere l’unità premendo lo stesso pulsante.
     
  5. Aprire FaroArm Manager, fare riferimento a FaroArm e ricollegare l'i-Probe, se era stato rimosso.
    Deleting corrupted i-probe_Fig.5.jpg
  6. Poiché tutta la cronologia della compensazione delle sonde è stata cancellata, è necessario compensare tutte le sonde prima dell'uso.
    Deleting corrupted i-probe_Fig.6.jpg